Specifications
of LCP 4100.1 |
| Control |
digital |
| Maximum operating pressure |
35 MPa |
| Pump system |
Two sapphire pistons (4mm) link parallel, pistons work in opposite cycle |
| Precision of pressure measurement |
± 2% |
| Flow rate setting |
0,01 - 9,99 ml/min |
| Method of setting |
by step 0,01 ml/min |
| Precision of set flow rate (1 ml/min, 12 MPa H2O) |
± 2% |
| Adjustable lower pressure limit |
0 - 39,9 MPa (by step 0,1MPa) |
| Adjustable upper pressure limit |
0 - 40 MPa (by step 0,1MPa) |
| Material of pumping head |
stainless steel - option PEEK |
| Data entry |
16 element keyboard |
| Display |
2 lines 40 characters |
| Power requirements |
230 V ± 10%, 50 Hz |
| Power input |
50 VA |
| Dimensions (W x H x D) |
260 x 160 x 460 mm |
| Weight |
12 kg |
Specifications
of LCP 4100.2(stainless), 4100.3(PEEK) |
| Control |
digital |
| Maximum operating pressure |
LCP 4100.2 35 MPa
LCP 4100.3 20 MPa |
| Pump system |
Two sapphire pistons (4mm) link parallel, pistons work in opposite cycle |
| Precision of pressure measurement |
± 2% |
| Flow rate setting |
0,01 - 9,99 ml/min |
| Method of setting |
by step 0,01 ml/min |
| Precision of set flow rate (1 ml/min, 12 MPa H2O) |
± 2% |
| Adjustable lower pressure limit |
0 - 39,9 MPa (by step 0,1MPa) |
| Adjustable upper pressure limit |
0 - 40 MPa (by step 0,1MPa) |
| Material of pumping head |
stainless steel - LCP 4100.2 |
| PEEK - LCP 4100.3 |
| Mixing of gradient mobile faze |
to input of pump |
| Maximum flow in a gradient mode |
3 ml/min |
| Number of valves |
3 (A, B, C) |
| Maximum number of gradient steps |
50 |
| Concentration range B, C |
0 - 100% in 1% steps, component A is automaticly fill in 100% |
| Duration of a single gradient step |
1s - 999 min 59s |
| Maximum number of stored files (programmes) |
100 |
| Data entry |
16 element keyboard |
| Display |
2 lines 40 characters |
| Power requirements |
230 V ± 10%, 50 Hz |
| Power input |
50 VA |
| Dimensions (W x H x D) |
260 x 160 x 460 mm |
| Weight |
12 kg |